2308-900-01 Rev. A / July 2008
Return to
Section Topics
11-7
Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Section 11: SCPI Tables
:LOW <NRf>
Specify integration time (in sec) for low pulse
measurements (33.33E-06 to 0.8333) - accounts
neither for internal trigger delay nor user delay.
Setting applies when SYNC:STATe is ON.
3.333E-05
:LOW?
Query low integration time when SYNC:STATe
is ON.
:AVERage <NRf>
Set integration time in seconds for
measurement of pulse AVERage time -accounts
neither for internal trigger delay nor user delay.
Setting applies when SYNC:STATe is ON.
3.333E-05
:AVERage?
Query average integration time when
SYNC:STATe is ON.
:DIGitize <NRf>
Specify integration time (in sec) for digitizing or
burst measurements; 33.33e-6 to 0.8333. Setting
applies when SYNC:STATe is OFF.
3.333e-5
:DIGitize?
Query integration time for digitizing or burst
measurements when SYNC:STATe is OFF.
:STEP <b>
To perform a series of step measurements, set to
ON.
OFF
:STEP?
Query state for performing step measurements.
:UP <NRf>
Specify number of UP steps <0-20> (max is for
both up and down combined).
1
:UP?
Query number of UP steps specified.
:DOWN <NRf>
Specify number of DOWN steps <0-20> (max is
for both up and down combined).
1
:DOWN?
Query number of DOWN steps specified.
:TIME <NRf>
Specify step integration time for UP plus
DOWN steps: 33.3µsec–100msec.
200
μ
sec
:TIME?
Query step integration time specified for UP
plus DOWN steps.
:TimeOUT <NRf>
Step TimeOUT (other than the first step):
2msec–200msec.
2ms
:TimeOUT?
Query step TimeOUT setting for all steps but
the first step.
:INITial <NRf>
TimeOUT for the first step: 10msec–60secs.
2sec
:INITial?
Query TimeOUT setting for the first step.
:DELay <NRf>
Step delay: 0msec–100msec (in 10msec
steps).
0
:SEQuence
<NRf, NRf>
Define up to 19 delay sequence values using
the index and the time value (e.g. "1,1e-3" sets the
first delay to 1ms). Ranges: 1-19 (parameter 1) and
0 to 100e-3 (parameter 2).
0
Table 11-4
SENSe command summary (refer to
Table 2-3
,
Table 3-2
, and
Table 4-2
)
Command
Description
Default
parameter SCPI
*
No query or short form exists.
Summary of Contents for 2308
Page 2: ......
Page 12: ...This page left blank intentionally...
Page 32: ...This page left blank intentionally...
Page 106: ...This page left blank intentionally...
Page 140: ...This page left blank intentionally...
Page 208: ...This page left blank intentionally...
Page 220: ...This page left blank intentionally...