Model 2308 Portable Device Battery/Charger Simulator User’s Manual
Section 13: Calibration
Viewing calibration date and count
Viewing date and count from the front panel
Follow the steps below to view the calibration date and count from the front panel:
1. Press
the
MENU
key then choose
CALIBRATE UNIT
, and press
ENTER
. The instrument
will display the last date calibrated:
CALIBRATE UNIT
LAST ON 07/01/2008
2. Press
the
▲
key. The instrument will display the number of times it was calibrated:
CALIBRATE UNIT
TIMES= 1
3. Press
MENU
to return to the menu structure.
Acquiring date and count by remote
Use the
CAL:PROT:DATE?
and
CAL:PROT:COUN?
queries to determine the calibration date
and count respectively (see
Miscellaneous commands
).