Analog output
There are two analog output terminals: a 0.5/5 A output terminal for use when the battery channel
is programmed for either the 5 A or the 500 mA range and a 5/50 mA output terminal for use when
the battery channel is programmed for either the 50 mA or the 5 mA range. Connect to the
appropriate analog output terminal and the Source - terminal.
The analog outputs provide a voltage output based on the measured current as follows:
• 0.5/5 A Output referenced to Source - : 1 A/V (each volt out represents 1 A)
• 5/50 mA Output referenced to Source - : 10 mA/V (each volt out represents 10 mA)
Each terminal's output impedance is nominally 1000 Ohms.
Figure 2-8
Analog output Connections
Source -
5/50 mA
0.5/5 A
Oscilloscope