F-2
Contrast
Join
Type
Scale
Math
F Source*
F Points*
F Window*
Status
Off
Add
Sub
Mult
FFT
CH1
CH2
CH3
CH4
1024
4096
Rect**
Hanning
Flat Top
DISPLAY
Overview
•
Power spectrum
Indicates the signal power
of each frequency component in dBm.
dBm is referenced to
1mW and the equivalent voltage (0.316V peak/0.224Vrms) at a
50
Ω
load.
•
Display of time axis (YT/XY) waveform or frequency axis (FFT) waveform
The display is switched to either the time axis (YT/XY) waveform or frequency axis
(FFT) waveform of the selected trace (CH1 through CH4).
•
Window
3 types of windows (Rectangular, Hanning, Flat Top) can be selected.
•
Number of data points for computation
The number of data points for FFT computation can be selected (1024, 4096).
•
Cursor measurement
The dBm value and frequency at the cross point with the Time (frequency) cursor can
be measured (dBm at frequency).
FFT Menu Hierarchy
* :
Displayed when FFT is
selected with Math.
** : Abbreviation of Rectangular.