AED mode
90
Step 4 - Deliver shock
If the need for shock is detected, the shock symbol will blink and the device will ask
the user to press the shock button again.
Press “SHOCK” button again.
The shock will be delivered.
ATTENTION: the user must not touch the patient or conductive surfaces
in contact with him/her during shock delivery, under risk of suffering a
powerful electric discharge.
NOTE: The energy delivered is pre-adjusted. The user cannot alter this
protocol. For adult electrodes: 1st shock is 150 J and the following are
200 J; for child electrodes, all shocks are 50 J.
If clinical scans show that defibrillation is not recommended, the CardioMax will
announce: “TREATMENT NOT RECOMMENDED”.
Check if there was no movement of the patient during the analysis. If so, restart
the process. If not, remove pads and start the CPR (cardiopulmonary resuscitation)
procedure. Details on the next section.
Press shock
button!