User manual |
Specifications and safety
199
Input dynamic range and
offset differential voltage:
The equipment complies with 50.102.2 standard IEC
60601-2-27.
Tachycardia alarm time:
206 bpm (1 mV): 5 seconds.
206 bpm (0.5 mV): 5 seconds.
206 bpm (2 mV): 5 seconds.
195 bpm (2 mV): 5 seconds.
195 bpm (1 mV): 5 seconds.
195 bpm (4 mV): 5 seconds.
ST segment:
YES (check availability).
ECG electrodes (adult or
child):
COMPOSITION: Adhesive conductor with hydrogel and
silver sensor (Ag/AgCl), latex free.
USE INSTRUCTIONS: 1) Shave hairs, clean the spot and
dry with gauze. 2) Before removing the protective film,
attach the ECG wire to the electrode. 3) Remove the film
from the electrode and put it in place by pressing firmly.
PRECAUTIONS: Do not open the packaging before
use. Do not apply the electrodes on the nipples, bony
protuberance, cuts, lesions, scars, only on intact skin. Store
at temperatures between 15°C and 40°C. Non-sterile single-
use product. Discard after use.
ECG gel:
It is recommended the use of a gel suitable for medical
procedure, indicated for use in electrocardiography,
defibrillators and similar. Used as a contact medium for
the transmission of electrical impulses by means of an
electroconductor. Biocompatible. After each use, clean the
accessories according to cleaning recommendations.
NIBP - Non-Invasive Arterial Pressure
Technique:
Oscillometric.
Adult range:
Systolic: 40 to 260 mmHg.
Mean: 26 to 220 mmHg.
Diastolic: 20 to 200 mmHg.