Sensitivity is the equipment’s percent ability to correctly identify a shockable rhythm.
Specificity is the equipment’s percent ability to correctly identify a nonshockable
rhythm.
A
=
True positive.
B
=
False positive.
C
=
False negative.
D
=
True negative.
A true positive (A) is the equipment’s ability, in measurement units, to
correctly
identify a
shockable
rhythm.
A false positive (B) is the equipment’s ability, in measurement units, to
wrongly
recognize a
shockable
rhythm.
A false negative (C) is the equipment’s ability, in measurement units, to
wrongly
recognize a
nonshockable
rhythm.
A true positive (D) is the equipment’s ability, in measurement units, to
correctly
recognize a
nonshockable
rhythm.
Values measured with the AED using the specified database:
VF/VT
Nonshockable
rhythms
Shock INDICATED
329
23
NO shock INDICATED
10
454
Sensivity =
97.05%
Specificity =
95,18%