User manual |
Alarms and limits
68
To turn the alarm sound on again, press the PAUSE AUDIO button for 1
second.
The sound of the parameters of alarm can be turned on and off individually in the
"Alarm" menu and in parameter menus.
Physiological alarm test
To perform the alarm test do the following steps:
1 - Turn on the equipment, with the electrodes properly connected. Verify the BPM
value on the equipment screen and that there are no alarms.
2 - Through the e-Jog Control key on the panel, navigate to: PARAMETER
CONFIGURATION > ECG CONFIGURATION > ALARM. With this same key, select
one of the limits and then change its value so that the initially measured value is
outside the range. Then go back to the main screen.
3 - Having done this, the alarm should sound. If it does not, the alarm is probably
defective.
4
–
Repeat this procedure for the other modules, remembering to test with their
respective sensor and parameters.
5 - With the indication of an active technical alarm, press the button PAUSE AUDIO
for 1s and check on screen the suspended alarm indication for all parameters.
Wait for 60s and the alarm will activate. On the screen, the suspended alarm signal
disappears, indicating the return of the alarm sound.
The time the alarm stays turned off is 60 s.
To turn the alarm sound on again, press the PAUSE AUDIO button for 1
second.
The sound of the parameters of alarm can be turned on and off individually in the
"Alarm" menu and in parameter menus.