RTC - Real Time Check
General
This function allows you to program up to three different times for the device to carry out
a self-diagnosis of defibrillation, battery level, connected paddles, and verification of the
device's connection to the power outlet. The results of the diagnosis are sent to the computer
that has the RTC system software installed on it and is within reach of the network
.
When the RTC Central module is not acquired, the equipment will not have communication
with PC central, but it will be enable on the menu (details on page 142).
RTC Operation on the CardioMax 8 with transmitter
The configurations for the RTC function can be accessed through the menu CONFIGU-
RATION > GENERAL CONFIGURATIONS > RTC (see chapter Screen and Operation).
The configurations screen of the RTC function allows you to define the number of
automatic daily tests to be performed, configurable in values from 1 to 3 tests. It is
also possible to configure the time for each of these daily tests, carry out a manual
test, or check the status of the last test carried out.
Daily tests quantity
3
06:00
12:00
20:00
First Schedule
Second Schedule
Third Schedule
Perform test manually
Last test status
RTC
back
exit
1. Number of automatic daily tests. It can be
configured with values of 1 to 3 tests daily.
2. Time for performing the first daily test.
Configured by the user.
3. Time for performing the second daily test.
Configured by the user. This item will be
disabled when the number of daily tests is 1.
4. Time for performing the third daily test.
Configured by the user. This item will be
disabled when the number of daily tests is
less than 3.
5. Carry out a manual test, without the need for
prior configuration.
6. Displays the data from the last test
performed.