Alarms and limits
67
Technical alarms settings
When turned on, CardioMax returns to the last limits and setting made by the user.
In order to change the active technical alarms, the user must access the "Settings >
Technical alarm" menu.
Through the menu, it is possible to enable or disable the sound indications of
technical alarms of ECG, SpO
2
, NIBP, EtCO
2
and Printer. When the sound is
enabled and there is a technical alarm situation in the parameter or the equipment,
the equipment will send beeps at intervals of approximately 6 seconds.
TECHNICAL ALARMS SOUND
ECG
Off
On
[Off ]
SPO2
Off
On
[Off ]
PANI
Off
On
[Off ]
EtCO2
Off
On
[Off ]
PRINTER
Off
On
[Off ]
Technical alarm test
To perform the alarm test do the following steps:
1 - Turn the equipment on, without cables or sensors connected. It should occur
indication of technical alarm (text messages in Infocenter) disconnected sensor.
2 - After confirming the visual indication, connect the sensor correctly. If the
indication disappears, the alarm is working properly.
Otherwise, change the sensor and repeat the test. If the result is the same, the
alarm will probably be defective.
3 - Repeat this procedure for the other modules, remembering to test with their
respective sensor and parameters.
4 - With the indication of an active technical alarm, press the button PAUSE AUDIO
for 1 s and check on screen the suspended alarm indication for all parameters.
Wait for 60 s and the alarm will activate. On the screen, the suspended alarm signal
disappears, indicating the return of the alarm sound.
The time the alarm stays turned off is 60 s.