WARNING: CardioMax should not be used too close to or over other
equipment. If this is necessary, it is recommended that the equipment or
system be observed to verify the normal operation in the configuration in
which it will be used.
NOTICE: When removing the equipment from its package, carefully verify if
there is any abnormality or visible damage in the device or its accessories,
caused by impact or mishandling during transportation. In case of
irregularities, please contact Instramed.
NOTICE: Disposable accessories and any other components must be
disposed of according to the norms of hospital waste disposal.
WARNING: Do not modify this equipment without authorization from
Instramed.
Adverse or side effects
Superficial burns may occur on the patient’s skin in the area in contact with the
electrodes. To minimize the effect of the disposable paddles, apply them directly after
removal from the protection envelope and attach them firmly to the patient’s skin.
The skin must be dry, or electric current leakage may occur, increasing the burn’s
area and reducing the efficiency of the treatment.
Possibility of reddish skin at the defibrillation electrodes exposition place, due to the
high voltage applied.
Possibility of superficial skin burns. To minimize the effect, in the case of adhesive
pads, apply them immediately after removal of the protective envelope and securely
attach to the patient’s skin. The patient’s skin must be dry.
Possibility of reduction of treatment efficiency. The patient’s skin must be dry,
otherwise the electric discharge may leak.
Do not apply conductive gel outside the area of contact with the pads to avoid
electrical leakage.
Possibility of erythematous skin at the pacemaker electrodes exposition place,
especially during prolonged use.
Possibility of discomfort and pain, with the use of high levels of pacemaker current.
In case of high levels of current, it is recommended that the patient is anesthetized or
unconscious.
Possibility of eczematous skin due to the use of non-biocompatible ECG electrodes
(non compliance with the ISO 10993 standard). Always use accessories with technical
recommendation described in this User’s Guide.
It is recommended for cases of resuscitation maneuvers of long duration, the use of a
gauze between the skin and the CPR MAESTRO.