Alarms and limits
60
Defibrillator
ALARM
PRIORITY
COLOR
POSSIBLE CAUSE
Connect pads
MEDIUM PRIORITY
YELLOW
Pads are not connected.
Ready load
HIGH PRIORITY
RED
It indicates that the load is complete and ready to
be delivered.
Defibrillator
failure
HIGH PRIORITY
RED
Defibrillator module failure.
INFORMATIVE MESSAGES
Adhesive pads -
CardioMax detected Adhesive pads connected.
Adult pads -
CardioMax detected Adult pads connected.
Children pads -
CardioMax detected paddles connected, but the adult electrode is disconnected. The same
message is also exhibited when CardioMax detects Children Adhesive pads.
Charging –
The load is not yet complete.
Shock delivered -
The energy was delivered to the patient.
Shock disarmed –
The energy was canceled by the operator.
Shock disarmed (Automatic internal discharge) -
The energy was automatically canceled because the shock
button was not pressed before the internal discharge time selected by the operator.
Shock disarmed (Contact Bad) –
The energy was canceled because the impedance of the patient is not
satisfying the applicable shock conditions.
Bad Contact –
Patient impedance measurement is not satisfying the applicable shock conditions.
Pacemaker
ALARM
PRIORITY
COLOR
POSSIBLE CAUSE
Connect
adhesive pads
MEDIUM PRIORITY
YELLOW
Adhesive paddles are disconnected.
INFORMATIVE MESSAGES
PM: Pause -
The outbreak of pacemaker pulses is paused.
PM: On -
The outbreak of pacemaker pulses is on.
Mode: Asynch –
The pacemaker is in asynchronous mode.
Mode: Dem –
The pacemaker is in demand mode.