Testing specifications for cabinet interface immunity to RF wire communications equipment
The Ion PRO was designed to provide Basic Security with RF equipment by the following table:
Testing
frequency (MHz) Band
a
(MHz)
Service
a
Modulation
b
Maximum
power (W)
Distance (m)
Immunity test
level (V/m)
385
380-390
TETRA 400
Pulse
modulation
b
18 Hz
1.8
0.3
27
450
430-470
GMRS 460
FRS 460
FM
c
deviation
of ± 5 kHz
Sinusoidal
1 kHz
2
0.3
28
710
704-787
Band LTE 13, 17
Pulse
modulation
b
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM 800/900
TETRA 800
iDEN 920
CDMA 850
Band LTE 5
Pulse
modulation
b
18 Hz
2
0.3
28
870
930
1720
1700-1990
GSM 1800
CDMA 1900
GSM 1900
DECT
Band LTE 1, 3, 4, 25
UMTS
Modulação de
pulso
b
217 Hz
2
0.3
28
1845
1970
2450
2400-2570
Bluetooth
WLAN, 802.11 b/g/n
RFID 2450
Band LTE 7
Pulse
modulation
b
217 Hz
2
0.3
28
5240
5100-5800
WLAN 802.11 a/n
Pulse
modulation
b
217 Hz
0.2
0.3
9
5500
5785
NOTE: If necessary, to reach the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the EM EQUIPMENT or
EM SYSTEM may be reduced to 1 m. The test distance of 1 m is allowed by ABNT NBR IEC 61000-4-3.
a
For some services, only the terminal’s transmission frequencies are included.
b
The carrier must be modulated using a 50% service cycle square wave signal.
c
As an alternative to FM modulation, 50% to 18 Hz pulse modulation can be used because, although it does not represent actual
modulation, this would be the worst case.