Monitor mode - Capnography
125
4 - Calibration
The equipment is calibrated using a known gas sample, thus configuring its
measuring curve.
IMPORTANT: Calibration must be performed when the equipment shows
a message requiring this procedure during initialization. Calibration must
be performed by QUALIFIED TECHNICAL PERSONNEL.
5 - EtCO
2
alarm
The “BELL” icon indicates audio “OFF”, “ON” or “PAUSED”. Alarm limits
configuration “MINIMUM” and “MAXIMUM”.
Is it possible to adjust the EtCO
2
minimum alarm to levels between 18 and 96
mmHg, with a 3 mmHg interval when in “Adult” mode and a 1 mmHg interval when
in “Neo” mode. It is possible to adjust the EtCO
2
maximum alarm to levels between
21 and 99 mmHg, with a 3 mmHg interval when in “Adult” mode and a 1 mmHg
interval when in “Neo” mode.
6 - Back/Exit
“BACK” to Configuration menu or “EXIT” to monitoring screen.