User manual |
Specifications and safety
188
Charge indicators:
Audio indication of equipment being charged, audio
indication of charge completed, LED on external pads and
charge level indicated on display.
Maximum charge time in
maximum energy:
< 6 s with 90% to 100% of the minimum specified voltage.
< 6 s with a full charge.
< 13 s from equipment initialization.
External pads size:
Adult = 10.3 cm x 8.5 cm (contact area = 81.9 cm²).
Children = 4.5 cm x 4 cm (contact area = 18 cm²).
.
Pads cable length:
2 meters.
The adhesive pads provided by Instramed have a
biocompatibility certificate in accordance with ISO 10993.
Cardioversion:
< 60 ms after the QRS peak.
Maximum output voltage:
2000 V.
Maximum output electric
current:
70 A (25 Ω).
AED mode
Waveform:
Truncated exponential biphasic pulse. Waveform
parameters adjusted according to the patient’s impedance.
Shock delivery:
By means of multifunctional pads (adhesive) or defibrillation
pads.
Defibrillation scales:
Adult: 150 and 200 J.
Child: 50 J.
Adult/children’s selection:
Automatic due to the size of the pads.
Charge command:
Automatic after identifying an arrhythmia.
Shock command:
Frontal panel button, "shock".