User manual |
Specifications and safety
197
Refractory period:
340 ms (between 30 to 80 ppm), 240 ms (between 90 to
180 ppm).
Maximum output voltage:
350 V.
ECG
Input:
3 or 5 lead ECG cable.
10 lead ECG cable (optional).
External pads.
Multifunctional pads.
Supports up to 12 simultaneous derivations when equipped
with optional cable.
Electrode error:
NO ELECTRODES and a trace line will appear on the
display if there is a disconnected cable or electrode.
Low ECG amplitude or
saturated ECG signal:
The “Searching for ECG Signal” message will appear on the
screen or printer (electrocardiograph mode) when the ECG
10 mm/mV amplitude is less than 2.4 mm peak-to peak (or
sensitivity proportional).
Multifunction pads error:
If an adhesive pad is disconnected, a trace line will appear
on the display.
Range:
Precision:
15 to 350 BPM.
± 1 BPM of 15 to 350 BPM.
Rejection in common mode:
Higher than 90 dB, measurement in compliance with the
AAMI standards for heart monitors (EC 13).
Sensitivity:
5, 10, 15, 20, 30 and 40 mm/mV.
CA line filter:
60 Hz or 50 Hz.
ECG response frequency:
Diagnostic mode (0.05 - 100 Hz).
Monitor mode (1 - 40 Hz).