![Instramed CardioMax User Manual Download Page 141](http://html.mh-extra.com/html/instramed/cardiomax/cardiomax_user-manual_2070096141.webp)
User manual |
RTC - Real Time Check
141
Automatic test
The automatic test can be programed through the menu
CONFIGURATION > GENERAL CONFIGURATIONS > RTC. To perform the
automatic test, the equipment should be turned off at the programmed time.
At the configured time, the CardioMax will automatically turn on and begin the test,
composed of the following steps:
• Test of the connection with the ECG board.
• Test of the defibrillator module.
• Check of the battery level.
• Check of the connection with the power outlet.
□
□
□
WARNING
RTC AUTO TEST
PRESS E-JOG TO CANCEL
Communication with the ECG
Defibrillation - 30J
Battery level
Connected to the main power
It is possible to interrupt the test with a click on the e-Jog button.