RTC - Real Time Check
144
It is not possible to interrupt the manual test. After completing the test, the
equipment will send the collected data to the RTC Control Center. At this time, it is
not possible to turn off the device. If the connection with the RTC Control Center
cannot be established, the device will cancel sending the data after 10 seconds.
After this procedure, the device will save the collected data and return to the mode
defined by the rotating switch on the CardioMax.
Status of the last test
To see the data from the last test performed, access CONFIGURATION >
GENERAL CONFIGURATIONS > RTC > STATUS OF THE LAST TEST. In this
window, the information regarding the last test is available, including possible
failures.
OK
Communication with the ECG
□
□
□
Defibrillation - 30J
Battery Level
Connected to the main power
98%
Failure in the communication with the Central
Last test status
To return to the RTC menu, just click on the "OK" button.
Error BIPS
When there is an error in the last RTC test, the equipment sends three "beeps" per
minute that can be remotely identified by the user.