EonStor S12F-R1432/G1433 Installation and Hardware Reference Manual
Hardware XOR
Yes
Real-time Clock
For event messages with time record and task
scheduling
A.3 Power
Supply
Specifications
Specifications
Nominal Power
530W with active PFC (S12F-R1432)
350W with active PFC (S12F-G1433)
DC Output
S12F-R1432
12.0V: 43A (Max.)
5.0V: 25A (Max.)
S12D-G1433
12.0V: 25A (Max.)
5.0V: 25A (Max.)
3.3V: 20A (Max.)
Input Frequency
47 to 63Hz
Input Voltage
AC 100-240 V ~ 10A-5A (S12F-R1432)
AC 100-240V ~ 6A-3A (S12F-G1433)
Power Factor
Correction
Yes
Hold-up Time
At least 20ms (S12F-R1432) / 16ms (S12F-
G1433) at 115/230VAC full load after a loss of
AC input
I
2
C
Through backplane to RAID controller(s)
Over-temperature
Protection
Auto shutdown when losing cooling or
elevated ambient temperature; over-voltage
protection is also available.
A.4
Cooling Module Specifications
Specifications
Max. Air Flow (each
module)
S12F-R1432
High speed: 90CFM
Low speed: 70.7CFM
S12F-G1433
High speed: 31.8CFM
Low speed: 20.1CFM
Input power
S12F-R1432
13.2W max
S12F-G1433
9 / 3.12W max.
Rated Voltage
DC 12V
Temperature
S12F-R1432
A-4