Multitron Pro - Incubation Shaker
Interferences
Page
162
of
185
2013-02-19
Temperature is not reaching
the desired value
Heating is not
switched on
Activate the temperature parameter Operator
Setpoint value is too
low
Increase setpoint value
Operator
Capillary thermostat
is set too low (must
be at least 10 °C
above the setpoint
value)
Set capillary thermostat to higher
temperature
Operator
Heating fan is not
working
Contact Infors representative
Qualified
technician
Door is not
completely closed
Close the door properly
Operator
Top-mounted
cooling:
cooling fins are
dusty
Contact Infors representative
Qualified
technician
Side-mounted
cooling:
cooling fins are
dusty
Clean cooling fins
Operator