9.
Open the door.
10.
Turn the thermostat down slowly until there is an audible click
at the actuation point of the thermostat.
11.
Set the thermostat 5 °C higher.
NOTICE!
The thermostat must be set a few °C higher than
the maximum operation temperature to allow a
continuous control.
The thermostat is now set. The maximum operating temperature
now corresponds to the specified setpoint shown in the display.
6.8 Setting the adjustable counterweight
The counterweight with a variable throw is equipped with two
detachable weights that can be flexibly mounted.
These weights are positioned according to the throw and the load.
The throw can be altered with an adjustable axle.
CAUTION!
Risk of material damage due to use of too high
shaking speed!
The incubation shaker may move due to vibrations
caused by too high shaking speeds and may cause
damage at the site.
Therefore:
– Never set speed value over given values of the
specifications in chapter „Technical Data“.
– Place equipment on an anti-slip matting
– Inform operator about throw settings