Multitron Pro - Incubation Shaker
Interferences
Page
164
of
185
2013-02-19
Illumination interference
Interference
Possible causes
Trouble shooting
By
Illumination (algae) is not
working
Illumination is not
switched on
Switch on the illumination
Operator
Illumination
parameter is not
activated
Activate the illumination parameter
in the control system
Operator
Lamp ballast is
defect
Contact Infors representative
Qualified
technician
Tube is defect
Contact Infors representative
Qualified
technician
Illumination (algae) is not
working
Control is defect
Contact Infors representative
Qualified
technician
Daylight illumination is not
working
Illumination
parameter is not
activated
Activating the illumination
parameter
Operator
Tubes have not
been turned on with
the switch inside the
chamber
Switch on
Operator
Lamp ballast is
defect
Contact Infors representative
Qualified
technician
Tube is defect
Replace tube
Operator
Starter is defect
Replace the starter
Operator
Daylight illumination is not
working
Control is defect
Contact Infors representative
Qualified
technician