Multitron Pro - Incubation Shaker
Operation
Page
124
of
185
2013-02-19
Inserting the tray
Procedure
1.
Place the tray on the glide rails of the door.
2.
Load the tray or handle the flasks.
Fig. 84
3.
Push the tray between the guide rails over the front stops until
it meets the stops at the back of the incubation chamber.
The tray must click into place behind the front stop.
Fig. 85
4.
Check whether the tray sits behind the front stop.
Closing the door lowers the tray ejectors and thus lowers the tray
into the two cones. At the same time, the locking hooks move
forwards to secure the tray to the table.
CAUTION!
Risk of material damage if the tray is not
locked.
If the tray is not secured to the table by the locking
mechanism, the tray can damage the chamber
when the table is moving.
Therefore:
– Only start the unit when the tray is securely
locked to the table.
5.