Multitron Pro - Incubation Shaker
Operation
Page
126
of
185
2013-02-19
Unlocking the tray manually
If there is a fault (power failure or mechanical problem) and the
door cannot be completely opened, the tray can also be removed
manually (by two people, if possible).
Procedure
1.
Unload the tray.
2.
Open the door as far as possible.
Fig. 87
3.
Person 1: Press both hooks of the locking mechanism
backwards.
4.
Person 2: Lift the tray out of the locking mechanism.
5.
Pull the tray over the front stop on the table.