Multitron Pro - Incubation Shaker
Operation
Page
128
of
185
2013-02-19
CAUTION!
Risk of material damage due to collisions
between the flasks and the walls of the
incubation chamber.
If the flasks extend beyond the edge of the tray,
they may be damaged or broken if they collide with
the wall of the unit.
Therefore:
– Always position culture flasks on the tray so that
they do not extend beyond the edge.
Procedure
1.
Open the door of the unit.
2.
Pull out the tray over the glide rails up to the stops.
3.
Load the tray evenly.
4.
Position any optional measuring sensors.
5.
Push the tray back all the way to the stop.
6.
Close the door.
NOTICE!
If the tray has a low load, it is advisable to place a
few flasks filled with water on the tray. This will help
the table to run more smoothly.
The incubator shaker is designed for an average
load of 12 kg.
The permissible tray loads are given in the chapter
"Technical Data".
NOTICE!
Over the short term, heavier or lighter loading
weights will not have an acute damaging effect as
a result of increased wear of the incubator shaker.
However, if heavier or lighter loads are to be used
at high speeds over the long term please contact
your INFORS representative.