Humidity interference
Interference
Possible causes
Trouble shooting
By
Water supply to the Direct
Steam Humidification is
interrupted; humidity
parameter has been
automatically deactivated.
Water reservoir is
empty
Fill water reservoir
Operator
Air pocket between
wet filter and
reservoir
Replace wet filter with a dry one
Operator
Valve is defect
Contact Infors representative
Qualified
technician
Transmission error
of the humidity
sensor or the
temperature sensor
Restart software (switch unit OFF
and then ON)
Operator
Steam generator is
defect
Contact Infors representative
Qualified
technician
Inlet pressure is too
high
Reduce inlet pressure (max. 0.3
bars)
Operator