8.2 Environmental protection
CAUTION!
Environmental danger by inappropriate
handling!
Inappropriate handling of environmentally
hazardous substances, especially where disposal
is involved, may lead to severe environmental
damage.
Therefore:
– Immediately take appropriate action if
environmentally hazardous substances are
accidentally released into the environment.
8.3 Maintenance plan
The required maintenance for reliable operation is described in the
following chapters.
Reduce the maintenance intervals in case increased abrasion is
detected during regular checks.
Contact the manufacturer for questions concerning maintenance.
For contact details, see page 2.
For proper operation the incubation shaker temporarily needs
following maintenance work:
Interval
Maintenance work
To be carried out
by
Before each use
Check that the chamber lights are working
Replace light bulbs if necessary
Operator
Option Direct Steam Humidification
Check integrity of hoses
Replace if necessary
Check water level in the tank
Top up if necessary
Operator
Option CO
2
gassing
Check gas supply has the correct settings
Adjust if necessary
Check integrity of hoses
Replace if necessary
Operator