Multitron Pro - Incubation Shaker
Operation
2013-02-19
Page
125
of
185
Fig. 86
6.
Check whether the locking hooks are firmly securing the tray
when the door is being closed
Withdrawing the tray
Procedure
1.
Stop the incubator shaker.
NOTICE!
The incubator shaker can be stopped either by
switching off the speed rpm or by half-opening the
door and waiting until the table has come to a
complete stop.
2.
Open the door to its full extent.
The safety mechanism automatically disengages the hooks when
the door is opened and lifts the tray out of the cones at the rear.
NOTICE!
If the tray cannot be pulled out, check whether the
door is completely open.
Press the door down to its full extent.
3.
Lift the tray by the handle.
4.
Pull the tray over the front stop.
5.
Pull the tray over the glide rails on the opened door.