4.8 Option cooling system
The incubator shaker is equipped with a cooling system so that
cultivation can be carried out in a defined climate.
The heat exchanger of all cooling systems is located behind the
rear wall of the incubation chamber.
The following cooling systems are available:
Top-mounted cooling system, 900 watts / 1200 watts
Fig. 30
The cooling unit is located in a housing mounted on top of the
incubator shaker.
It can cool up to three incubators.
The cooling unit has a separate power supply.
The mains switch is located on the left-hand side of the mounted
housing.
The cooling unit has a twin-circuit system with a coolant tank.
The coolant tank is located in the mounted housing.
Side-mounted cooling system, 380 watts
Fig. 31
The cooling unit is located in a housing mounted on the side of the
incubator shaker.
It can cool only one incubator.
The incubator shaker supplies power to the cooling unit.
The cooling unit is a closed single-circuit system.