6.7 Setting the capillary thermostat
The capillary thermostat of the incubation chamber has a default
setting of 70 °C. This is 5 °C above the maximum operating
temperature of the unit, e.g. 65 °C.
The absolute tolerance of the thermostat is 2.5 °C.
Procedure
1.
Set the thermostat to 70 °C.
2.
Shut the door of the unit.
3.
Switch on the unit.
4.
Press option key to select the temperature parameter.
The temperature parameter lights up.
5.
Press Set key.
Setpoint lights up.
6.
Press the PLUS key until the maximum operating temperature
(e.g. 50 °C) is set.
7.
Press the ON/OFF key to switch on the temperature control
system.
8.
Wait until the upper display has reached the setpoint value
(e.g. 50 °C).