Multitron Pro - Incubation Shaker
Interferences
2013-02-19
Page
161
of
185
No value on temperature
display
Pt100 sensor or
sensor cable is
defect
Contact Infors representative
Qualified
technician
Temperature in flask with
mobile Pt100 is too high
(> 65 °C)
Defect measuring
electronics
Contact Infors representative
Qualified
technician
Temperature in flask with
mobile Pt100 is too low
(< 0 °C)
Defect measuring
electronics
Contact Infors representative
Qualified
technician
No value on temperature
display
Pt100 sensor or
sensor cable is
defect
Contact Infors representative
Qualified
technician
Temperature too high.
No error message
Cooling is not
switched on
Switch on cooling
Operator
No power supply to
cooling system
Check fuse of cooling system,
replace if necessary
Operator
Connect unit to power supply
Operator