Multitron Pro - Incubation Shaker
Interferences
2013-02-19
Page
165
of
185
9.4 Interference Illumination broken tube
CAUTION!
Damage to health due to inhalation or skin
absorption of mercury.
Therefore:
– Avoid skin contact
– Ventilate the room for 30 minutes
– Wear safety gloves
– Collect shards and dust with an adhesive tape
– Pick up pieces of fluorescent tube with a
cleaning cloth
– Seal contaminated clothing, gloves, pieces of
tube and cleaning equipment so that they are
airtight
– Ensure they are disposed of correctly
If a fluorescent tube has broken, proceed as follows:
Procedure
1.
Switch off the unit.
2.
Disconnect the unit from the power supply.
3.
Remove fluorescent tube and shards out of the incubation
chamber.
For optional illumination (fluorescent tubes)
4.
Remove the fluorescent tube out of the lamp socket.
5.
Fit a new fluorescent tube.
For optional algae illumination (installed through the housing wall)
The defect fluorescent tube must be exchanged by a qualified
technician (in-house technician, manufacturer's service team).