Maintenance
2013-02-19
Page
151
of
185
CAUTION!
Risk of material damage due to the use of
unsuitable cleaning agents.
Aggressive cleaning agents, solvents and scouring
cleaning aids (hard sponges, brushes) may scratch
and damage the surfaces and have a negative
impact on the function of the unit.
Therefore:
– Use only soft cloths.
– Use only mild household cleaners.
8.11.1 Cleaning the housing
Wipe the housing regularly with a soft cloth and a mild household
cleaner (e.g. washing-up liquid, neutral cleaner). If necessary,
disinfect with a standard disinfectant.
8.11.2 Cleaning the chamber
Wipe the housing regularly with a soft cloth and a mild household
cleaner (e.g. washing-up liquid, neutral cleaner). If necessary,
disinfect with a standard disinfectant.
8.11.3 Cleaning the floor pan
When cleaning the floor pan after a glass breakage or if culture
liquids have escaped, the table can be released from the
counterweight and tilted upwards.
Clean the floor pan as follows:
Procedure
1
.
Open the door.
2
.
Switch off the unit and disconnect it from the power supply.
Fig. 108
3
.
Remove the Allen screws (4) from the drive hub of the table
using the supplied Allen key.
4
.
Tilt the table upwards.
5
.
Remove any foreign objects and debris by hand.