Multitron Pro - Incubation Shaker
Operation
Page
136
of
185
2013-02-19
Set the CO
2
gassing system as follows:
Procedure
1.
Press option key to select parameter CO
2
.
Parameter sign and unit light up.
2.
Press SET key.
3.
Press PLUS or MINUS key to change setpoint.
4.
Press ON/OFF key to switch on parameter.
Display shows current value (e.g. 5.0%).
Using the safety curtain
The safety curtain is attached to the ceiling of the incubation
chamber with four snap fasteners.
NOTICE!
The label „Outside“ must face outwards.
Fig. 90
Detaching the safety curtain
Undo each snap fastener with a gentle tug.
Attaching the safety curtain
Gently press the two halves of the snap fastener together, one by
one.