7.10 Alarm messages
A message or an alarm is usually triggered by an operating error
with the incubator shaker.
To call the users attention according to a problem, the warning sign
and ALARM icon always lights up.
The message or alarm is automatically cancelled when the setpoint
value is reached.
The following messages are displayed:
Alarm message
Possible causes
Trouble shooting
By
Temperature is too low
Deviation > 1 °C
Door is not fully closed. Close the door properly.
Operator
Fan is not operating.
Inform Infors representative.
Qualified
technician
Capillary thermostat is
set too low.
Set capillary thermostat to higher
temperature.
Operator
Temperature is too high
Deviation > 1 °C
Cooling is not switched
on.
Switch on cooling.
Operator
Fan is not operating
Inform Infors representative.
Qualified
technician
Setpoint is below the
technically feasible
value.
Increase setpoint value.
Operator
Speed is too low
Deviation > 10 rpm
Tray is too heavily
loaded
Load tray properly
Operator
Wear on drive belt
Contact Infors representative
Qualified
technician
Speed is too high
Deviation > 10 rpm
Control system is
defect
Contact Infors representative
Qualified
technician
CO
2
concentration is too low
Deviation > 1%
CO
2
valve is closed or
blocked
Contact Infors representative
Qualified
technician
CO
2
cylinder is empty
Install new CO
2
cylinder
Operator
Inlet pressure or flow
rate is too low
Increase inlet pressure or flow
rate
Operator