![horiba HU-200TB SS-120 Instruction Manual Download Page 56](http://html1.mh-extra.com/html/horiba/hu-200tb-ss-120/hu-200tb-ss-120_instruction-manual_2162225056.webp)
Measurement
48
Cleaning operation
Automatic cleaning is operated in accordance with the following conditions.
"CLn"
Cleaning
type
Internal
cleaning
cycle
Cleaning:
○
Standby:
-
External
contact input
ON:
○
OFF:
-
Cleaning
operation
*1
ON:
◎
OFF:
-
"CLn"
Contact
output
*2
ON:
●
OFF:
-
"HoLd"
Contact
output
*3
ON:
●○
OFF:
-
Transmission
output
"yES"
"nor"
○
○
◎
●
●
*4
○
-
◎
●
●
-
○
-
-
○
-
-
-
-
-
Measured
value
"or"
○
○
◎
●
●
*4
○
-
◎
●
●
-
○
◎
●
●
-
-
-
-
-
Measured
value
"And"
○
○
◎
●
●
*4
○
-
-
-
-
Measured
value
-
○
-
-
-
-
-
-
-
-
"trg"
No setting
ON for 2 s
◎
●
●
*4
-
-
-
-
Measured
value
"no"
No setting
No setting
○
-
-
○
*4
-
-
-
-
Measured
value
*1 Cleaning operation runs in accordance with the specified method when "Setting the automatic cleaner" of the
Setup menu is set to "yES".
*2 The "CLn" contact output is enabled in the HOLD state during cleaning and after cleaning.
*3
●
indicates a case where the "Hold" contact output is enabled in the HOLD state during cleaning and after
cleaning.
○
indicates a case where the "Hold" contact output is enabled when the external contact input is ON.
*4 The transmission output is enabled in accordance with the transmission output hold settings in the Setup menu.