Measurement
34
Decimal-point
placement
(display range)
"Auto"
"1000"
"100.0"
"10.00"
"Auto"
Select the automatic changeover or
fixed decimal point.
With a turbidity value, Auto moves the
placement of the decimal point
automatically.For the fixed zero point,
select the optimal placement of the
decimal point against the turbidity of a
sample.
Set the auxiliary
display part (sub-
display part)
"r1"
"r2"
"non"
"non"
The setting defines what to display on
the sub-display part when returning
from the Hold mode to the Meas.
mode.
"non" does not display anything.
"r1" displays an alarm value or
assignment for Relay 1.
"r2" displays an alarm value or
assignment for Relay 2.
Set out of
measurement
range hidden
(range cut)
"yES"
"no"
"yES"
z
"yES": The range cut is enabled.
For the fixed zero point, a
displayed value is fixed at the
maximum value and blinks when it
exceeds the maximum value of
each display range.
A displayed value is fixed at 10.00,
100.0 or 1000 within each decimal
point range.
For the automatic shifting of the
decimal point, a displayed value is
fixed at 1000 and blinks when it
exceeds 1000. However, the
displayed value of Polystyrene
Latex is fixed at 100 degrees and
blinks.
z
"no": The range cut is disabled.
For the fixed zero point, a
displayed value is fixed at any
value when it exceeds the
maximum value plus 10% of each
display range.
A displayed value is fixed at 11.00,
110.0 or 1100 within each decimal
point range.
For the automatic shifting of the
decimal point, a displayed value is
fixed at 1100 and blinks when it
exceeds 1100.
However, the displayed value of
Polystyrene Latex is fixed at 11.00,
125.0 or 125 degrees and blinks.
Screen 1
Screen 2
Description
Setting range
Initial
value
Remarks