Installation
17
Note
The maximum load resistance for transmission output is 900
Ω
.
J
Connecting the contact input terminal
z
The applicable electric wire is of 0.14 mm
2
to 2.5 mm
2
(AWG26 to AWG14).
z
For a cable for contact input, use a twist pair shielded wire.
z
If lighting might strike, install an arrester along the signal line.
Note
The contact input impedance should be within 100
Ω
.
When contact input 1 is short-circuited, the EXT1 indicator on the display part will light up.
(+) (-)
OUT1
(transmission output)
RS-485
SENSOR
EXT.IN
OUT1
RS-485
SENSOR
EXT.IN
Contact input 1
Open-circuit voltage
24 V DC
Short-circuit current
12 mA maximum
OUT1