Measurement
33
Table 2 List of display items in the Setup menu
Screen 1
Screen 2
Description
Setting range
Initial
value
Remarks
Set the moving
average count
(damping factor)
1 to 120
times
20
Increase dF to reduce the variations
in measured values. Decrease dF to
speed up the response.
Set auto return
"yES"
"no"
"yES"
This setting defines whether or not to
select auto return from the Hold mode
to the Meas. mode.
Selecting "yES" enables auto return.
Selecting "no" disables auto return. It
will be necessary to return to the
Meas. mode manually.
Set auto return
time
1 to 24 (
h
)
2
This setting returns from the Hold
mode to the Meas. mode after the set
auto return time has elapsed since
the last key operation.In the Hold
mode, a transmission signal remains
in the HOLD state.To release the
HOLD state, it is necessary to return
to the Meas. mode.
Select a turbidity
standard
substance
"SiO"
"For"
"PSL"
"For"
Select any standard turbidity
substances from among Kaolin (SiO),
Formazin (For), and Polystyrene
Latex (PSL).
Select the
turbidity units and
the SS units
"PtnA"
"Ptnb"
"PtnA"
For Kaolin, turbidity and mg/L can be
further selected as the unit of
measure.
Pattern A TURB
Pattern b mg/L (SS)