Measurement
37
Set the
ransmission
output range
(Range)
"10.00"
"100.0"
"1000"
10.00
Selects a transmission output range.
A zero point and a full-scale point can
be arbitrarily selected within the
Range.
Set the
transmission
output zero point
0.00 to 10.00
0.0 to 100.0
0 to 1000
0.00
The transmission output zero point
can be freely selected within the
Range.
Define a value equivalent to 4 mA.
Set the
transmission
output span point
0.00 to 10.00
0.0 to 100.0
0 to 1000
10.00
The transmission output span point
can be freely selected within the
Range.
Define a value equivalent to 20 mA.
Set the
transmission hold
of transmission
output (Current
Hold)
"HoLd"
"PrES"
"HoLd"
Define a transmission output value for
the time of transmission output hold.
z
"Hold": The last value is output
while being held.
z
"PrES": The preset value is output.
Set the arbitrary
value (when the
transmission hold
is set to "PrES")
0.00 to 10.00
0.0 to 100.0
0 to 1000
10.00
The HOLD state will be maintained at
the pre-set value.
Set the
transmission
output burnout
"non"
"out.4"
"out.20"
"non"
Specify the burnout of transmission
output for the occurrence of an
abnormality during measurement.
z
out.4: Outputs 3.8 mA at the time
of burnout.
z
out.20: Outputs 21 mA at the time
of burnout.
z
Non: There is no burnout setting.
Screen 1
Screen 2
Description
Setting range
Initial
value
Remarks