Measurement
35
Relay 1
Relay 2
Set contact output
types ("r1" and
"r2")
"non"
"AL"
"HoLd"
"CLn"
"non"
Specify a contact output type.
z
"non": No output
z
"AL": Operates at alarm contact
point.
z
"Hold": Outputs during
transmission output hold.
z
"CLn": Outputs a trigger during
cleaning or after cleaning.
Set a control
method (when the
contact output
type is "AL")
"Hi"
"Lo"
"Hi"
Specify whether to control by upper
limit operation or lower limit
operation.
z
"Hi": control by the upper limit
operation
z
"Low": control by the lower limit
operation
Select the
decimal point
place (range) of a
numerical values
for alarm settings
(when the contact
output type is set
to "AL".)
10.00
100.0
1000
10.00
Select the optimal place of the
decimal point (range) for the alarm
settings.
Set alarm value
(when the contact
output kind is
"AL")
0.00 to 10.00
0.0 to 100.0
0 to 1000
10.00
Define an alarm value.
Press the ENT key to make changes
possible, make adjustments with the
▲
/
▼
key, and then press the ENT
key to confirm the changes made.
Set contact output
delay time (when
the contact output
type is "AL" or
"Hold").
0 to 600
0
Define the delay time (in seconds)
from the time an alarm is raised until
a contact is output.
If the alarm contact point is released,
an alarm will be immediately released
without delay.
Set a cleaning
output type (when
the contact output
type is "CLn")
"CLn"
"trg"
"CLn"
Specify a cleaning output type.
z
"CLn": Outputs a cleaning output
type in the HOLD state holding
during and after cleaning.
z
"trg": Performs the contact output
for 5 s within 2 s after cleaning
ends.
Screen 1
Screen 2
Description
Setting range
Initial
value
Remarks