Measurement
36
Set the FAIL
output when an
error is judged as
an abnormality by
the self-check
function.
"yES"
"no"
"yES"
Specify whether or not to trigger FAIL
output when an error is judged as an
abnormality by the self-check
function.
(E-80 to E-85)
Set the FAIL
output at the full-
scale over of a
measured value
"yES"
"no"
"no"
Specify whether or not to output FAIL
when a measured value goes beyond
the full scale.
If FAIL output is enabled,
z
If range cut is enabled,
1000 degrees (Kaolin, Formazin)
100 degrees (Polystyrene Latex)
z
If range cut is disabled,
1100 degrees (Kaolin, Formazin)
125 degrees (Polystyrene Latex)
FAIL is output when a measured
value exceeds the above limit.
Set FAIL output
delay time
0 to 600
0
Define the delay time (in seconds)
from the time FAIL is triggered until
the contact point output is performed.
If a FAIL contact point is released,
FAIL is immediately released.
Screen 1
Screen 2
Description
Setting range
Initial
value
Remarks