Operation
26
Q
Initial setup
The converter is shipped with initial settings from the factory.
If it is necessary to make any change to the initial settings, make the settings in accordance
with the instructions given in “ Entering the Setup menu ” (page 32).
For the initial settings, see the table below.
To return to the Meas. mode, hold down the ESC key until the HOLD indicator turns OFF.
Setting
menu items
Screen
display
Description
Initial
value
Reference
page
Parameter
Setting
"PArA"
"dF"
Set the moving average count (damping factor).
20
33
"A.rEt"
Set auto return.
"yES"
33
"rEt.t"
Set auto return time.
2
33
Sensor settings
"SEnSor"
"tyPE"
Select a turbidity standard substance (Type).
"For"
33
"Unit"
For Kaolin, select the turbidity units and the SS units (mg/L).
"PtnA"
33
Display settings
"diSP"
"d.rng"
Select the automatic shift of the decimal point and the fixed zero
point (10.00 100.0 1000).
"Auto"
34
"S.diSP" Set the auxiliary display part (sub-display part).
"non"
34
"rng.C"
Set out of measurement range hidden (range cut).
"yES"
34
Relay output
settings
Setting
"r1"
"r2"
"1.tyPE" Set contact output types ("r1" and "r2").
"non"
35
"1.HorL" Set a control method (when the contact output type is "AL").
"Hi"
35
"1.rng"
Specify a decimal place for alarm value settings (range selection).
10.00
35
"1.SEt"
Set alarm value (when the contact output kind is "AL" ).
10.00
35
"1.dt"
Set contact output delay time (when the contact output type is
"AL" or "Hold").
0
35
"1.Cln.t"
Set a cleaning output type (when the contact output type is
"CLn").
CLn
35
FAIL output setting
"FAiL"
"F.SC"
Set the FAiL output for abnormality detection by self-check.
"yES"
36
"F.oF"
Set the FAIL output for the case where a measured value
exceeds the full-scale value.
"no"
36
"F.dt"
Set FAIL output delay time (in seconds).
0
36
Transmission
output setting
"Curr"
"C.rng"
Set a full-scale range of transmission output (Current Range).
10.00
37
"rng.0"
Set a zero point for transmission output (arbitrarily
programmable).
0.00
37
"rng.S"
Set a span point for transmission output (arbitrarily
programmable).
10.00
37
"C.HoLd" Set the transmission hold for transmission output (Current Hold).
"HoLd"
37
"PrES"
Set a transmission hold value (when transmission hold is "PrES").
10.00
37
"b.out"
Set the burnout for transmission output (Burnout).
"non"
37
Cleaning operation
"CLn"
"CLn"
Set whether or not an automatic cleaner will be used.
"no"
38
"tyPE"
Set a cleaning operation type (when the use of an automatic
cleaner is set to "yES").
"or"
38
"CyCLE"
Set a cleaning cycle for the internal timer (when the use of an
automatic cleaner is set to "yES").
"3.0"
38
"CLn.t"
Set cleaning time (when the use of an automatic cleaner is set to
"yES").
"20"
38
"Hold.t"
Set the extended time for transmission output hold after the end
of cleaning.
"120"
38
Setting for RS485
"rS"
"Addr"
Set RS-485 address.
0
38