Measurement
29
Meas. mode
Moving average setup screen
*Setup menu
If “non” is selected from the "diSP" setting
items, nothing is displayed in the sub-display
part after returning to the Meas. mode.
Controlled value setup
menu
Cleaning check screen
Calibration data check screen
Setting
change
Establishes
the value
Cancels
the value
Alarm settings can be changed during
measurement.
The upper or lower limits of motions cannot
be switched.
The delay time of an alarm contact point can
be changed.
When a numerical value is adjusted with the
▲
/
▼
key and the ENT key is pressed, the
changes will be confirmed.
If the use of the automatic cleaner is selected, the countdown
of the cleaning cycle timer can be checked.
If the ENT key is pressed in the Test mode, the set time
cleaning will start.
The calibration history can be checked.
The value of an amount of transmitted light obtained during
the latest zero calibration is compared with the factory-
configured value, and is expressed in %. This screen shows
the number of days elapsed since the last zero calibration.
The moving average count can be changed during measurement.
When a numerical value is adjusted with the
▲
/
▼
key and the
ENT key is pressed, the changes will be confirmed.