7–8
Rapid fFN 10Q
System User Manual
Section 7 — Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
12
Invalid liquid
control test
result.
Internal Controls
Analyzer QC:
Fail Cassette
QC: Pass
Refer to Error/Invalid Code table.
Turn analyzer off and back on to reinitialize the system.
Run menu 3-Daily QC. If the Daily QC result is Pass, retest
the control on a new cassette. If both pass, no further
troubleshooting is required.
If the Daily QC result is Fail, call Technical Support.
Internal Controls
Analyzer QC:
Pass
Cassette QC:
Fail
Verify that 200 uL of control was dispensed.
Review the Rapid fFN Control Kit directional insert to ensure
the correct procedure was followed. Repeat the test on
a new cassette. If control result is acceptable, no further
troubleshooting is required.
Do not bump or jar the analyzer or the cassette during the
test.
If control is invalid a second time, call Technical Support.