7–1
Rapid fFN 10Q
System User Manual
Section 7 — Troubleshooting
SECTION 7 — TROUBLESHOOTING
GENERAL INFORMATION
The Rapid fFN 10Q Analyzer software is designed for easy troubleshooting. Always heed the beep tones and
follow the display screen prompts to obtain the best performance from your system. The following table lists
potential problems, sources of trouble, and recommended solutions. Call Hologic, Inc. Technical Support for
any questions related to the performance of your Rapid fFN 10Q System.
ITEM
PROBLEM
SOURCE
SOLUTION
1
Analyzer display
screen is blank.
Analyzer Power
Cord and/or
Adapter
Ensure analyzer power cord is fi rmly connected to analyzer.
Ensure analyzer power adapter is plugged into a grounded
AC electrical outlet.
ON/OFF switch
Ensure analyzer ON/OFF switch is in ON position.
2
Error Code is
displayed when
analyzer is fi rst
turned on.
Analyzer
Turn analyzer off and back on to reinitialize the system.
If the Error Code persists, refer to Error/Invalid Code table.