3–16
Rapid fFN 10Q
System User Manual
Section 3 — General Operating/Testing Instructions
Insert cassette and press ENTER.
The analyzer will check that a Rapid fFN 10Q Cassette is properly inserted.
The instrument will beep repeatedly and the display will read ADD SAMPLE
AND IMMEDIATELY PRESS ENTER.
Add 200 uL of control sample and press ENTER.
The analyzer will begin a 7-minute incubation countdown.
Following incubation, the analyzer will begin analysis of the cassette.
When testing is complete, the system will display and print the result.
Note:
Both a Level 1 and a Level 2 control sample must be run. Repeat
the test with the other control sample and a fresh cassette.
If the Liquid Control result does not pass, refer to Section 7, Troubleshooting,
Item 11.
INSERT CASSETTE
PRESS ENTER
TO CONTINUE
CHECKING CASSETTE
DO NOT REMOVE
CASSETTE
ADD SAMPLE
AND IMMEDIATELY
PRESS ENTER
Rapid fFN 10Q
03:00 PM 10/26/13
LEVEL 1: M1023
fFN: 31 ng/mL
TEST IN PROCESS
ANALYZING
DO NOT REMOVE
CASSETTE