4–18
Rapid fFN 10Q
System User Manual
Section 4 —
Software Functions — LIQUID CONTROLS
The analyzer then prompts the user to insert the cassette and press
ENTER.
This message will be displayed if a cassette is not inserted. Press ENTER
to return to the previous screen.
The analyzer will check that a Rapid fFN 10Q Cassette is properly
inserted.
A 30-second timer starts during which time this message fl ashes and the
analyzer beeps. Add 200 uL of control sample and
immediately
press
ENTER.
INSERT CASSETTE
PRESS ENTER
TO CONTINUE
CASSETTE NOT
INSERTED
ENTER-CONTINUE
ADD SAMPLE
AND IMMEDIATELY
PRESS ENTER
CHECKING CASSETTE
DO NOT REMOVE
CASSETTE