2–1
Rapid fFN 10Q
System User Manual
Section 2 — Installation
SECTION 2 — INSTALLATION
GENERAL
This section provides detailed installation instructions for the Rapid fFN 10Q
System. Follow installation steps
carefully to insure proper installation and operation.
ENVIRONMENTAL FACTORS
The Rapid fFN 10Q System has been designed to be safe under the following conditions:
• Indoor use;
• Altitudes up to 2000 m;
• Maximum relative humidity of 80% for temperatures up to 30°C;
• Pollution Degree II, in accordance with IEC 61010-1 Category II. The Rapid fFN 10Q System is for use only
in an offi ce or a clean laboratory environment.
• The operating temperature should be held relatively constant. The optimum operating temperature is 15° to
30°C (64° to 86°F). Before operating, allow the instrument to equilibrate to room temperature.
Place the instrument away from direct sunlight and away from locations subject to extreme temperature
variations (e.g., near open windows, ovens, hot plates, radiators, etc).