7–10
Rapid fFN 10Q
System User Manual
Section 7 — Troubleshooting
ERROR/INVALID CODES
ERROR CODE
DEFINITION
TROUBLESHOOTING PROCEDURE
003, 004, 007, 020,
024, 100, 103, 104,
107
Internal Controls
Cassette QC Failure
Refer to Troubleshooting Items 12 and 13.
142A
Dark Count Out of Range
Turn analyzer off and back on to reinitialize the system.
Retest Daily QC.
If Daily QC passes, continue with normal operation. If Daily QC
fails, contact Technical Support.
142B
Sensor Error
Turn analyzer off and back on to reinitialize the system.
Retest Daily QC.
If Daily QC passes, continue with normal operation. If Daily QC
fails, contact Technical Support.
622
Possible Voltage Problem
Turn analyzer off and back on to reinitialize the system.
Retest Daily QC.
If Daily QC passes, continue with normal operation. If Daily QC
fails, contact Technical Support.
720
721
Possible Motor Problem
Turn analyzer off and back on to reinitialize the system.
Retest Daily QC.
If Daily QC passes, continue with normal operation. If Daily QC
fails, contact Technical Support.
If analyzer errors persist or recur, call Technical Support.